M
Michael LaneNov 27, 2025
What to do if the density displayed is deviating from the actual density in YOKOGAWA Rotamass Supreme?
- CCarlos MaddenNov 27, 2025
If the density displayed on your YOKOGAWA Measuring Instruments is deviating from the actual density, consider the following potential causes: * The density unit, highest, and lowest value for density may be set incorrectly. Match the settings of the flow meter and reading system, and check the LRV and URV process parameters. * A fixed density might be selected. Check whether the Val sel parameter is set correctly. If a fixed value is selected, ensure that the Fix val parameter is set correctly. Otherwise, set the Val sel parameter to Meas val. * The analog output trim might have been performed incorrectly, requiring you to correctly perform trimming. * There might be no electrical grounding. Ground the transmitter and sensor, and check the correct connection of the connecting cab...











