EasyManua.ls Logo

YOKOGAWA UT150 - Page 44

YOKOGAWA UT150
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-14
<Toc> <Ind>
TI 05C01E02-01E 1st Edition : Oct. 31, 2001-00
The following operating procedure describes an example of setting the “deviation high
and low limit” (setpoint: 7) for the alarm-1 type, and “5°C“ for the alarm 1 hysteresis.
Step 1:
Bring the operating display into view.
UT130
Display example
UT150/UT152/UT155
Display example
Step 3:
Press the key several times to display
the parameter "CTL"(control mode).
Set "-1" to enter the setup parameter
setting display. But if "LOC" = 1 or 2 is
already set, the parameter value can not
be changed by setting "LOC" = -1 only. To
change the parameter value, set "LOC" = 0
at first (for disabling key lock), then set
"LOC" = -1 once again
NOTE
Step 5:
Press the key to display "-1".
Step 6:
Press the key once.
Flashes during change.
Flashes during change.
Step 2:
Press the key for 3 seconds or more
to display the parameter "A1".
The parameter "A1" appears only for the
controller with the "/AL" or" /HBA" option.
The parameter "CTL" appears for the
controller without the "/AL" or" /HBA" option,
and in this case, the alarm function is not
available.
Step 7:
Press the key several times to display
the parameter "AL1" (alarm 1 type).
The parameter "AL1" appears only for the
controller with the "/AL" or" /HBA" option.
To the next page
To the next page
Step 4
(for UT130 only)
:
Press the or key once to display
the setpoint.
<Operating Procedure>

Table of Contents

Other manuals for YOKOGAWA UT150

Related product manuals