6-1
IM WT1801-01EN
6 Holding Measured Values and Performing Single
Measurements
Holding Measured Values (HOLD)
The hold operation stops data measurement and display operations for each data update interval and holds the
display of all measurement function data. The D/A output values, the values in the data lists printed by the built-
in printer, and the communication values all reflect the held values.
For information about how holding works during integration, see the sections on holding and releasing
integration.
Click here.
Single Measurement (SINGLE)
While the display is held, the signal is measured once at the data update interval, and then the display is re-held.
If you press SINGLE when the display is not held, measurement restarts from that point.
Using an External Signal to Hold Measured Values and Perform Single Measurements
(Option)
On models with the 20-channel D/A output option, you can use the remote control feature to hold measured
values and perform single measurements. For details about the remote control feature, see appendix 4.6 in the
getting started guide, IM WT1801-03EN.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT1800 and is the answer not in the manual?

YOKOGAWA WT1800 Specifications

General IconGeneral
ModelWT1800
Basic Power Accuracy±0.1% of reading
Frequency RangeDC, 0.1 Hz to 1 MHz
Harmonic AnalysisYes, up to 500th order
Voltage Range1.5 V to 1000 V
Current Range0.5 A to 50 A
Display8.4-inch LCD
InterfacesUSB, Ethernet, GP-IB
Sampling Rate2 MS/s
Dimensions426 mm x 177 mm x 400 mm