Mechanical cleaning/disinfection (validated process)
Phase Step T[˚C/˚F] t[min] Waterquality Comment
I Pre-rinse < 40 /104(cold) 3 T-W
Emptying
II Cleaning 55/131 10 T-W neodisher®
MediClean Dental
Emptying
III Rinse < 40/104 2 T-W
Emptying
IV Rinse < 30/86(cold) 1 VE-W
Emptying
V Thermo-
disinfection
90/194 5 VE-W
Emptying
VI Drying > 90 /104 30
Seite 39/69
Page 39/47JMA-Optic
T-W: Drinking water
VE-W: Fully demineralized water (demineralized, microbiologically
at least drinking water quality)
Validated with: neodisher® MediClean Dental in a Miele PG 8535
Workflow:
Carefully place the product in a suitable rinsing basket.Make sure that the product does
not touch any other instruments/instrument parts (no rinsing shadow).
Start a tested program with the following characteristics:
Sufficient cleaning effect.
Final rinse with distilled or fully demineralized water.
Thermal disinfection with deionized water: 5 minutes at 90 - 95 C , A0 >=3000.
Sufficient product drying (no visible liquid residues).
Check and pack the components immediately after removal,
see chapter: 8.4./8.5.Inspection and packaging, p. 39/40.
8.4. Inspection
All components must be checked for the following in each case after cleaning/disinfection:
Damage to the product, e.g. cracks, chipping.
Damage to the surface
Deformations on the product or parts of the product.
If the product still has visible contamination after the end of the program, repeat the pre-cleaning
and machine cleaning/disinfection. Damaged components must not be used any further.