EasyManuals Logo

Zeiss GOM Scan 1 User Manual

Zeiss GOM Scan 1
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
Fig. 9: The image shows the first calibration panel positions (1– 12) in the
measuring volume.
From calibration step 13, the calibration panel is only tilted and rotated in
the center of the measuring volume. The new holding device for the cali‐
bration panel facilitates the positioning due to the dierent images for the
calibration panel.
If you have not set the automatic exposure time, readjust the exposure
time for each new image.
Follow the further instructions on the GOM software until the procedure is
complete.
8.4.4 Calibration Result
At the end of the calibration procedure, the software displays the calibra‐
tion result. The green icon
indicates a good calibration result. The red
icon
indicates an insucient calibration result.
If the calibration result is insucient, repeat the calibration and check the
sensor settings and the correct camera lenses. Ensure that the cameras
and lenses are mounted tight.
If only the projector calibration is insucient, check if the projector lens
and the lens settings are correct and ensure that the lens is mounted tight.
Calibrate the Sensor
0000002121_003_EN_01-09-2021 Page 20 (24)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zeiss GOM Scan 1 and is the answer not in the manual?

Zeiss GOM Scan 1 Specifications

General IconGeneral
BrandZeiss
ModelGOM Scan 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals