86
000000Ć1150Ć839 IOLMaster 11.02.2004
Line voltage 100 ... 240 V AC ("10 %)
Frequency 50 ... 60 Hz
Power consumtion 90 VA
Power connection Power cord with power plug with earthing contact
Enclosure Protection
Protection Class
Device type
IP 20
I
B (according to EN 60601Ć1)
Dimensions
Bench space
Height
390 mm x 300 mm
max. 610 mm (headrest)
Mass approx. 18 kg
Ambient conditions for intended use Temperature: )10 ... )35 °C
Relative humidity: 30 ... 75 %
Air pressure: 800 ... 1060 hPa
Storage conditions Temperature: -10 ... )55 °C
Relative humidity: 10 ... 95 %
Air pressure: 700 ... 1060 hPa
Shipment conditions (in original packing) Temperature: -40 ... )70 °C
Relative humidity: 10 ... 95 %
Air pressure: 500 ... 1060 hPa
Measuring range
Axial length
Range
Resolution of result display
Corneal radius
Range
Resolution of result display
Anterior chamber depth
Range
Resolution of result display
White To White (optional)
Range
Resolution of result display
14 ... 40 mm
0.01 mm
5 ... 10 mm (33 ... 67 D)
0.01 mm
1.5 ... 6.5 mm
0.01 mm
8 ... 16 mm
0.1 mm
Technical Data