Measurement information
Comparison results of IOLMaster versus conventional measurements at the human eye*:
Axial length**
Corneal radius***
Anterior chamber depth**
Mean values of Differences Standard Deviations
-0.01 mm $0.19 mm
-0.01 mm $0.05 mm
0.07 mm $0.18 mm
IOLMaster measuring reproducibility*:
Axial length
Corneal radius
Anterior chamber depth
Regarding the human eye regarding the artificial model eye
standard deviation**** standard deviation
$0.03 mm $0.01 mm
$0.02 mm $0.02 mm
$0.10 mm $0.10 mm
* Data
on
File
** Compared
to
high
precision
ultrasound
immersion
device
*** Compared
to
a
handheld
keratometer
**** Standard
deviation
determined
on
basic
of
x
i
"
calculated
single
standard
deviations
Technical Data