What does Error No. 452 mean on Zeiss Medical Equipment?
J
Jimmy ThompsonSep 7, 2025
If your Zeiss Medical Equipment displays Error No. 452, it means the Illumination unit cannot be detected due to a CAN-BUS error. Replace the instrument to resolve this.
J
Julia BoydSep 6, 2025
What to do if Zeiss Medical Equipment shows Error No. 92?
Z
zhughesSep 6, 2025
If your Zeiss Medical Equipment displays Error No. 92, it indicates a Power Module EEPROM error. To resolve this, replace the Power Module.
R
raycollinsSep 5, 2025
What does Error No. 80 mean on Zeiss Medical Equipment?
D
dennis16Sep 6, 2025
If your Zeiss Medical Equipment displays Error No. 80, it indicates an Imaging Module EEPROM error related to ID exchange. Replace the Power Module to fix this.
C
Carrie DuncanSep 5, 2025
What causes Error No. 79 on Zeiss Medical Equipment?
J
James FosterSep 5, 2025
If your Zeiss Medical Equipment displays Error No. 79, it indicates a Power Module EEPROM writing error. To resolve this, replace the Power Module.
R
Robert RothSep 4, 2025
How to fix Error No. 78 on Zeiss Medical Equipment?
J
Joshua WernerSep 4, 2025
If your Zeiss Medical Equipment displays Error No. 78, it indicates a Power Module EEPROM reading error. The suggested solution is to replace the Power Module.
R
rmaxwellSep 3, 2025
What should I do if my Zeiss Medical Equipment shows Error No. 77?
T
Thomas MitchellSep 3, 2025
If your Zeiss Medical Equipment displays Error No. 77, it indicates a Power Module EEPROM error related to detecting the last "triples". Replace the Power Module to resolve this issue.
J
jefferywatkinsAug 25, 2025
What to do if I get Error No. 90 on my Zeiss Medical Equipment?
A
Aaron WhiteAug 25, 2025
If you're encountering a USB-Bus error (Error No. 90) on your Zeiss Medical Equipment, first check or replace the connected USB devices. If the problem persists, the Power Module may need to be replaced.
K
Kathryn FuentesSep 2, 2025
Why does Zeiss Medical Equipment show Error No. 834?
F
Francisco JonesSep 2, 2025
If your Zeiss Medical Equipment shows Error No. 834, it means the equipment cannot reach the desired magnification position. Check all microscope cable connections. If the error continues, replace the Microscope.
B
Barbara WalkerSep 1, 2025
How to troubleshoot Error No. 448 on Zeiss Medical Equipment?
C
Christopher BurtonSep 1, 2025
If your Zeiss Medical Equipment displays Error No. 448, it indicates an Operator Panel error. Check all connections at the Operator Panel. If the issue continues, replace the Operator Panel or the instrument.
B
Beth IbarraSep 1, 2025
What to do if Zeiss Medical Equipment shows Error No. 447?
L
Laura SheltonSep 1, 2025
If your Zeiss Medical Equipment displays Error No. 447, it indicates an Operator Panel error. Check all connections at the Operator Panel. If the issue continues, replace the Operator Panel or the instrument.