Maintenance
13.3 Functional test
13
43
Function test
1. Ensure that all persons present in the treatment room are wearing
protective glasses.
2. Conduct a performance test (see chapter 7).
3. Press the emergency-stop-button. The device must switch off immediately.
Return the button into its original position by turning the knob in the
direction shown on by the arrows.