+20% +15% +0% -25%
Crystalloid
Colloid +0% -5% -20%
Crystalloid
Colloid
Whole blood
Packed RBCs
+25% +20% +10% -25%
+15% +10% -15%
+10% -10% -35%
-5% -25%
Cartridge IV Fluid
Peripheral IV Catheter (Ga)
equivalent central venous catheter lumen
14 12 16 14 18 16 20 18 22 20
Crystalloid/
Colloid Cartridge
(PN 0040-0050)
Blood
Cartridge
(PN 0040-0051)
Recommended Do not use
* Effect of catheter size less at lower rates
May trip occlusion alarm; use larger
catheter if this occurs
Effect of Catheter Size on Max Flow Rate
% difference, actual v. set flow rate @ 6*