It is very important to transfer the code data after an event as you have a limited amount of time (30 min to 3
days per hospital configuration) before the R Series will erase all content. Follow these steps for WiFi Transfer
or Compact Flash Card Transfer:
WiFi Transfer
Monitor Mode———Report Data——Transfer Mode———Report to WiFI
Compact Flash Card
Monitor Mode———Report Data——Transfer Mode———Report to Card
Test Log
The R Series can hold the last 999 Tests performed (manual or automatic) and can be reviewed/printed out for
verification at anytime. Test Log will never become “full” as once the memory reaches 999 and a new test is
entered, the oldest test will be deleted to make room. Follow the steps to review the Test Log:
Monitor Mode———Report Data——Test Log

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB