R Series ALS In-service Guide
OVERVIEW
Color Coded Operating System:
Review with the Mode Selector dial.
o RED for Defibrillation
o GREEN for Pacing
o GREY for Monitoring.
Power Indicators
Battery Indicator
o GREEN = Charged
o AMBER = Charging
o FLASHING = Problem (check battery)
o Indicator lights on battery show 30 min. increments
for a total of 4 hours of runtime.
o If orange light appears on “?”of battery, call BioMed
to reset.
A/C Power Indicator
o GREEN = Device is plugged in
o No light = Device is NOT plugged in
Test & Code Readiness Indicator
Every 24 hours, the device will do a 100 point test and will display a
Green check in window = Device is ready for use
Red x in window = Failed a part of the test. Device is NOT ready for use.
The device will display “Readiness Test Failed” on the screen and will
highlight what needs to be fixed. These are usually easily fixable problems
(No pads plugged in, no battery, no A/C power, etc).
After fixing any easily fixable problems, perform a manual test at 30J with
pads plugged in. “30J Test OK.” “Defib Pad Short.” and green check in
window indicates pass.
If the device reads “DEFIB FAILURE”, call BioMed.
Setup
A OneStep CPR Electrode should ALWAYS be connected to the therapy cable.
o Ensures “Code-Readiness”
o Allows the device to pass its Automated Daily Test.
Checks
Daily Check: 2 lights, a green check mark, the cap snugly on the back
Weekly Check: Follow the “Weekly Test” instructions in your Code Log books
Recorder Tray
Press button to open door
Insert the paper with the black arrow facing towards the back of the device
Pull a strip or two forward and let paper drop
Close the door on top of the strip.

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals