EasyManuals Logo

ZOLL R Series In-Service Guide

ZOLL R Series
5 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #3 background imageLoading...
Page #3 background image
DEFIBRILLATION (Red Zone)
Put rhythm simulator on “VT HI”
AED (Automated External Defibrillator)
1. Turn to DEFIB mode
2. Press ANALYZE
o Stand clear, analyzing
o If shockable rhythm, shock button will illuminate red
3. “All clear!”, then press SHOCK.
Auto Escalation
The device will increase the energy automatically
Energy will start at 120J, then to 150J, then to 200J for Adult Patients
Energy for Pediatrics starts at 50J, 70J, 85J
Manual Defibrillation
1. Turn to DEFIB mode
2. Press ENERGY SELECT arrows to change the energy level
3. Press CHARGE
4. “All clear”, then press SHOCK
CPR SENSOR
CPR Sensor
Is built right in to the pad
Enables CPR Feedback based on RATE, DEPTH, RECOIL
Idle Timer: Displayed on monitor after 3 seconds of inactivity.
Rate and Depth Numerical Values: Displayed once you start compressions.
Release Bar: Indicates release and recoil off the chest. Full Bar = Great!
Diamond/PPI (Perfusion Performance Index): Mimics perfusion pressure and will begin to fill up.
Goal
1. Depth: 2.0-2.4” for adults
2. Rate 100-120 CPM
3. Recoil: Release Bar needs to be full
Practice
Scenario 1 - Not doing fast enough compressions. Metronome sound will begin. Match the
metronome to get back to the right rate of 100cpm.
Scenario 2 - Not going deep enough.
o The device will say “Push harder.” to get you above 2 inches in depth.
o Once you have sufficient depth, the device will say “Good compressions.
Pediatric Feedback - Will only display Rate and Depth- you will not hear the “Push Harder”
voice prompt
———————————BLS USERS ARE NOW EXCUSED FROM TRAINING—————————

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB