6 APPENDIX B: ALTERNATE CONFIGURATIONS
Dual Unit Conguration
The haptic device may be congured in a dual unit setup. In this scenario, two units are connected upon beginning
the Smart Setup application.
1. With two devices connected, select the Touch Smart Setup desktop icon to open the Touch Smart Setup application.
2. When prompted, select both the Right and Left devices from the list.
3. Insert the stylus for both devices into each inkwell for calibration.
NOTE: Please refer to the Run 3D Systems Touch Smart Setup section for detailed calibration steps.
4. Perform the Sphere Simulation and Hollow Cube tests on each device to verify proper performance.
5. Select Save Conguration to complete the dual unit conguration process.