Ranger
™
Model 245
Blood/Fluid Warming System
For I.V. Use
3
Ranger
™
Model 247
Irrigation Fluid Warming System
Not for I.V. Use
1  Model 245 • Blood/Fluid Warming System
• For I.V. Use
1  Model 247 • Irrigation Fluid Warming System
• Not for I.V. Use
3  Modell 245 • Blut- und Infusionswärmesystem
• Für intravenöse Anwendungen
3  Modell 247 • Spülflüssigkeit-Erwärmungssystem
• Nicht zum intravenösen Gebrauch
2  Modèle 245 • Système de réchauffement de
sang/soluté • Pour perfusion I.V
2  Modèle 247 • Système de réchauffement de liquide
d’irrigation • Ne pas utiliser pour une perfusion I.V
Service Manual 2
Manuel d’entretien 23
Service-Handbuch 45

Other manuals for 3M Ranger 247

Questions and Answers

  • M
    Megan MasonSep 12, 2025
    What to do if 3M Ranger 247 Medical Equipment alphanumeric display reads “Er 4” or “Open.”?
    • P
      Pamela BakerSep 12, 2025
      If the alphanumeric display on your 3M Medical Equipment reads “Er 4” or “Open,” it indicates an open wire on the temperature sensor. Do not use the unit.
  • M
    markhillSep 9, 2025
    What to do if 3M Ranger 247 Medical Equipment warming unit displays low temperature alert?
    • W
      willisstefanieSep 9, 2025
      If the 3M Medical Equipment warming unit's alphanumeric display flashes a temperature of 33°C or below along with the word “LO,” and the alert indicator illuminates with a sound, the alert should stop when the temperature rises above 33°C. This under-temperature condition may be caused by very high flow using very cold fluid, or a defective heater/relay. If the alert continues, turn off the unit, unplug it, and discontinue use.
  • J
    Joan CookSep 8, 2025
    What to do if 3M Ranger 247 Medical Equipment alphanumeric display reads “Er 5” or “Open.”?
    • M
      Mark BeanSep 8, 2025
      If the alphanumeric display on your 3M Medical Equipment reads “Er 5” or “Open,” it indicates electrical interference. Remove the unit and refer to a biomedical technician.
  • M
    Melissa MontgomerySep 7, 2025
    How to remove cassette from 3M Ranger 247?
    • T
      thomas04Sep 7, 2025
      To remove the cassette from the 3M Medical Equipment, ensure that: * Fluid is drained from the cassette before sliding it out. * Fluids are no longer infusing. * The clamp is closed proximal to the warming cassette. Also, if the warming unit is below the patient level, raise it above the patient level to avoid excessive back pressure.
  • A
    amber80Aug 23, 2025
    What to do if the 3M Ranger 247 warming unit displays a high temperature alert?
    • J
      Jill GonzalezAug 23, 2025
      If the alert indicator illuminates and you hear alert sounds, and the display flashes a temperature of 48°C or higher along with the word “HI,” it indicates a temporary over-temperature condition. This can happen due to rapid changes in flow rates, starting the unit before inserting the warming cassette, or using fluids pre-warmed above 42°C. To resolve this, increase the flow to lower the temperature. The alerts will stop when the temperature drops below 48°C, and the unit will be ready for use. If this doesn't work, turn off and unplug the unit, stop the fluid infusion, and avoid pre-warming fluids before using the 3M Ranger irrigation fluid warming unit.
  • M
    Michele MorganSep 5, 2025
    Why nothing illuminates on the 3M Medical Equipment warming unit panel?
    • P
      pdavisSep 5, 2025
      If nothing illuminates on the warming unit panel of your 3M Medical Equipment, it could be due to several reasons: * The unit is not turned on. * The unit is not plugged in, or the power cord is not correctly plugged into an appropriate outlet. Ensure the power cord is plugged into the power entry module of the warming unit. * There may be a unit failure. Check the fuses.
  • D
    Daniel PerezSep 4, 2025
    What to do if 3M Ranger 247 alerts sound, alphanumeric display and alert indicator light go dark?
    • C
      christopher58Sep 4, 2025
      If the alert sounds, the alphanumeric display and alert indicator light go dark on your 3M Medical Equipment, this indicates a primary controller failure, and the unit will no longer operate. Turn the unit off and unplug it. Discontinue use of the unit and discard the warming set. Note that the alert will continue to sound if you do not unplug the unit.
  • C
    Christine GonzalezSep 3, 2025
    What to do if my 3M Ranger 247 alerts sound but unit has been turned off?
    • R
      Robin JonesSep 3, 2025
      If the 3M Medical Equipment continues to alert even after being turned off, it means that the independent backup safety system has been activated. Unplug the unit. Also, ensure the over-temperature screw is completely tightened. If it is missing, turn the unit off, unplug it.
  • J
    Jesus WhiteSep 1, 2025
    What to do if 3M Medical Equipment warming unit shows high temperature alert?
    • N
      nandersonSep 2, 2025
      If the 3M Medical Equipment warming unit's alphanumeric display flashes a temperature of 43°C (Model 245) or 48°C (Model 247) or higher, along with the word “HI,” and the alert indicator illuminates with a sound, consider these possibilities: * If an extreme change in flow rates occurred, open the flow to reduce the temperature. The alerts will stop when the display reads 41°C, and the unit will be ready to use. * If the unit reached the set point temperature before the cassette was inserted, the alerts will stop when the display reads 41°C, making the unit ready for use. * If fluids were prewarmed above 42°C, turn off the unit, discontinue the infusion, and avoid prewarming fluids before using the Ranger warming unit.
  • K
    Kevin RossSep 1, 2025
    What to do if the 3M Ranger 247 Medical Equipment warming unit displays a low temperature alert?
    • I
      ijohnsonSep 1, 2025
      If the alert indicator illuminates, alert sounds, and the display flashes a temperature of 33°C or below along with the word “LO,” it indicates an under-temperature condition. This can be caused by very high flow using very cold fluid, or a defective heater/relay. The alerts should stop when the temperature rises above 33°C. If alerts continue, turn off the unit, unplug it, and discontinue use. Call 3M Patient Warming technical support.

Summary

Technical Service and Support

Contact Information for Service

Details on contacting Arizant Healthcare for technical assistance and ordering.

System Setup and Functionality

Safety and Precautions

General Warnings and Cautions

Critical safety information, warnings, and cautions for operating the warming unit.

Product Description

Maintenance and Cleaning

Model 245 Preventative Maintenance

Model 245 Preventative Maintenance Procedure

Executing Model 245 Maintenance Checklist

Step-by-step instructions for performing the preventative maintenance on Model 245.

Model 245 Over-Temperature Testing

Testing Model 245 Alarm Points

Procedure for testing the Model 245 unit's over-temperature alarm set points.

Model 247 Preventative Maintenance

Model 247 Over-Temperature Testing

Testing Model 247 Alarm Points

Procedure for testing the Model 247 unit's over-temperature alarm set points.

Replacing the Power Fuse

Troubleshooting Guide

Troubleshooting Power and Alarm Issues

Solutions for common issues like no power, alarm sounds, or display problems.

Advanced Troubleshooting

Troubleshooting Unit Alarms and Errors

Addressing advanced issues such as unit alarms on power-up or error code display.

Technical Specifications

3M Ranger 247 Specifications

General IconGeneral