GX-A Series/GF-A Series Communication Manual 20 / 41
In addition to weighing data, other data can be added. Switch each internal setting on / off as
necessary.
Data number Internal setting
□ When the data memory function is used, the data number is output.
□ Consists of 6 characters (excluding the terminator).
ID number Internal setting
□ The ID number stored in the balance is output.
□ Consists of 13 characters (excluding the terminator).
Date Internal setting or
□ The date is output from the clock data of the balance.
□ The order of YYYY/MM/DD can be changed in settings.
□ Consists of 10 characters (excluding the terminator).
Time Internal setting or
□ The time is output from the clock data of the balance.
□ 24-hour format.
□ Consists of 8 characters (excluding the terminator).
7-4. Other data formats
N o . 0 0 1 CR LF
Data
number
Terminator
2 0 1 7 / 1 2 / 3 CR LF
Date Terminator
1
1 2 : 3 4 : 5 6 CR LF
Time Terminator
S A M P L E - 0 1 CR LF
ID number Terminator
2 3 - 4