EasyManua.ls Logo

ABB AWT210 - Cable Length; Cable Specification; Network Connection

ABB AWT210
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 AWT210 PROFIBUS PA 2-WIRE TRANSMITTER | COM/AWT210/PROFIBUS-EN REV. A
6 Installation
Cable length
Maximum cable length 1900 m (6233 ft)
Maximum spur length (1 device) 120 m (394 ft)
Maximum spur length (2 devices) 90 m (295 ft)
Maximum spur length (3 devices) 60 m (197 ft)
Maximum spur length (4 devices) 30 m (98 ft)
Cable specification
The cable lengths in Table 1 apply to cable to the following
specification:
 
Capacitance per unit length <30 pf/m
 
Core diameter 0.64 mm
Core cross section >0.34 mm
2
Network connection
PROFIBUS network connections are made to the terminal block
on the AWT210 Communication module.
WARNING – BODILY INJURY
Refer to the AWT210 Operating Instruction
(OI/AWT210-EN) before making electrical
connections.
NOTICE
Property damage
When connecting an AWT210 to a PROFIBUS-PA network:
Use cable that meets PROFIBUS specifications for
reliable communications
Ensure a PROFIBUS active terminator is fitted to
each end of each bus segment.
Route data lines clear of the source of any strong
electrical and magnetic fields.
Figure 2 PROFIBUS PA communication module connection overview
SW1:
Reset to
defaults
ON
OFF
SW2:
Write
protection
123456 87
AB
2
AB
1435
Sensor module connections
PROFIBUS PA module
connections
(connected internally)
Non IS installations
9 to 32 V DC
IS installations
9 to 24 V DC
Functional earth

Other manuals for ABB AWT210

Related product manuals