Mounting
SM/FEX300/FEX500/FM/CSA-EN FEX300, FEX500 EN - 19
4.7 Electrical connection
Electrical connection should be performed by following the information in the operating
instructions OI/FEX300/FEX500 or the commissioning instructions CI/ FEX300/FEX500.
According to NEC standards, the separate ground connection between flowmeter sensor and
transmitter can be established as follows:
1. Route the steel wire (1) in the signal cable under the clip.
2. Fit the steel wire (1) with protective tubing and wire end sleeve (2).
3. Connect to the grounding connection (3).
Dual compartment housing Single compartment housing
Fig. 12
G00760-01
1
2
3
1
2
3
1