Navigator 600 Series Analyzers
Silica and Phosphate Appendix B – Acyclic Parameter Mapping
10 IM/NAV6PBS Issue 1
Level sensors 1 182 Simple Unsigned8 1 R 0
0 = Disabled
1 = Enabled
Flow sensors 1 183 Simple Unsigned8 1 R 0
0 = Disabled
1 = Enabled
Update time 1 184 Simple Unsigned16 2 RW 15 1 to 20
Automatic schedule 1 185 Record 21 RW
Simple Unsigned8 1 0
0 = Off
1 to 7 = 1 to 7 days
Zero calibration frequency
Simple Unsigned16 2 Year Next zero date & time
Array Unsigned8 4
Month
Day
Hour
Minute
Simple Unsigned8 1
0 = Off
1 to 7 = 1 to 7 days
8 = 2 weeks
9 = 4 weeks
Secondary calibration frequency
Simple Unsigned16 2 Year Next secondary date
Array Unsigned8 4
Month
Day
Hour
Minute
Time from zero
Simple Unsigned8 1
0 = Off
1 to 7 = 1 to 7 days
8 = 2 weeks
9 = 4 weeks
Clean frequency
Simple Unsigned16 2 Year Next clean date
Array Unsigned8 4
Month
Day
Hour
Minute
Time from zero
1
Store definitions – refer to page 7
2
Access definitions – refer to page 7
Description Slot
Slot
Index
Object
Typ e
Data Type Bytes Store
1
Access
2
Default Valid Range Note
Table B.1 Data Structure (Sheet 4 of 38)