Silica and Phosphate Appendix B Acyclic Parameter Mapping
IM/NAV6PBS Issue 1 11
Clean durations 1 186 Array Unsigned16 4 RW 20, 20
20, 30, 60 mins
2, 4, 6, 8, 12 hours
Automatic and manual clean durations
Valve & pump status 1 187 Array Unsigned8 11 R
Streams 1 to 6.
Zero, secondary & clean valves.
Sample & reagent pumps.
Mode of operation 1 189 Simple Unsigned8 1 R 0
0 = Normal
1 = Demonstration
2 = Test
Archive definition 1 190 Record 6 RW
Simple Unsigned8 1 0
0 = Text
1 = Binary
Archive format
Array Unsigned8 3 0, 0, 0
Data, alarm event and audit logs
0 = Off
1 = On
Archive enable
Simple Unsigned8 1 2
0 = Off
1 = Hour
2 = Day
3 = Month
New file interval
Simple Unsigned8
10
0 = Off
1 = On
File wrapping
Filter type 1 191 Simple Unsigned8 1 RW 0
0 = Instantaneous
2 = Average
3 = Minimum only
4 = Maximum only
5 = Minimum & Maximum
Sample rate 1 192 Simple Unsigned32 4 RW 10000 1000 to 43200000 Sample rate in ms
1
Store definitions – refer to page 7
2
Access definitions – refer to page 7
Description Slot
Slot
Index
Object
Typ e
Data Type Bytes Store
1
Access
2
Default Valid Range Note
Table B.1 Data Structure (Sheet 5 of 38)