EasyManuals Logo

Abbott e i-STAT Alinity User Manual

Abbott e i-STAT Alinity
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
15
i-STAT
Alinity — Quick Reference Guide Art: 731848-01 Rev K Rev. Date: 02-Apr-2021
SECTION 5
DISINFECTING
i-STAT Alinity Instrument
Disinfection is recommended between each patient. When instrument is
dedicated to a single patient, disinfect at least once a day. The disinfection
process must begin
IMMEDIATELY
after the cleaning procedure is
complete. Standard precautions should be taken whenever working with
blood or blood products.
Remove a wipe from the
container and squeeze out
any excess uid before you
begin to wipe the surface.
Wipe all surfaces three times.
This includes front, back, sides,
top and bottom.
WIPEWIPE
3X 3X
WIPE WIPE
3X 3X
W
I
P
E
3
X
3
X
1 2
3 4
Wipe with gauze
until dry.
Allow all
surfaces to
remain wet
for 3 minutes.
SENSITIVE AREAS
Avoid forcing liquid into these areas:
i-STAT Alinity Instrument
Cartridge Port
10-Pin Connector under the camera
Gold contacts (2) on the outside of
the battery
A
B
C
C
B
A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott e i-STAT Alinity and is the answer not in the manual?

Abbott e i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modele i-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals