Page #18 background imageLoading...
Page #18 background image
16
i-STAT
Alinity — Quick Reference Guide Art: 731848-01 Rev K Rev. Date: 02-Apr-2021
SECTION 5
DISINFECTING
Base Station, Electronic Simulator and Printer
4
Wipe dry with gauze.
Remove a wipe from the
container and squeeze out
any excess uid before you
begin to wipe the surface.
Wipe the front as shown, ip products
on back side and repeat.
WIPE 3X
LEFT & RIGHT
1 2
WIPE 3X
UP & DOWN
SENSITIVE AREAS
Avoid forcing liquid
into these areas:
D
E
F
Base Station
10-Pin Connector
Gold contact pins (2)
USB Port
D
E
F
Electronic
Simulator
Area between
protective cap
retaining ring
and white
sensor area
G
WIPE
3
X
3
X
3X 3X
3
Allow all surfaces to
remain wet for 3 minutes.

Questions and Answers: