Export Error
If the «Export failed» window
appears, an unspecied error has
occurred during export of results.
This may have been caused by either
the storage medium or the Pima
Analyser.
Press or and remove the storage medium. Repeat export of results.
Should this fail a second me insert a dierent USB storage medium and
export results.
Power Warning
The «Warning Recharge baery»
window appears indicang that the
Pima Analyser's integrated baery is
low. This warning can appear at any
me, even during an acve analysis.
At this point the Pima Analyser is
fully funconal and the analysis can
be connued.
It is recommended to connect the Pima Analyser to A/C power before
starng another analysis.
Press or to connue.
9  ERROR WARNINGS & MESSAGES
2020-11-01 11:05
FAILURE
Export
failed
2020-11-01 11:05
WARNING
Recharge
battery