EasyManuals Logo

Accu-Chek ACTIVE Getting Started Guide

Accu-Chek ACTIVE
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #6 background imageLoading...
Page #6 background image
www.accu-chek.com.au
Accu-Chek Enquiry Line: 1800 251 816
Performa
Accu-Chek
®
Performa
Move into the shade or shield
the meter with your body,
for example.
Repeat the test.
Move to a place where the
ambient temperature is not
higher than +40 °C and wait
for the temperature of the
meter to adjust to this temperature.
Do not apply blood or control
solution to the test area until the
prompt apply drop is displayed.
Repeat the test.
Insert two new batteries.
Move to a place where the
ambient temperature is at
least +10 °C and wait for the
temperature of the meter to
adjust to this temperature.
Error Messages - Accu-Chek
®
Mobile
If the error message is displayed
after turning the meter on again:
Contact the Accu-Chek Enquiry Line on 1800 251 816.
10 11

Other manuals for Accu-Chek ACTIVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek ACTIVE and is the answer not in the manual?

Accu-Chek ACTIVE Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1 - 2 μL
Test Time5 seconds
Memory Capacity500 results
Display TypeLCD
Automatic CodingYes
Blood SampleCapillary whole blood
Dimensions97.8 x 46.8 x 19.1 mm
Battery Type1 x 3V lithium battery (type CR2032)
Battery LifeApproximately 1000 tests
Data TransferVia infrared (IR) to computer
Operating Conditions8-42°C, 10-85% RH