Performance Analysis
The performance data for the Accu-Chek Aviva Nano System (Accu-Chek Aviva Nano
Meter with Accu-Chek Aviva Test Strips) were obtained using capillary blood from
diabetic patients (method comparison, accuracy), venous blood (repeatability) and
control solution (reproducibility). The system is calibrated with venous blood
containing various levels of glucose. The reference values are obtained using the
hexokinase method. For method comparison, the results were compared with results
obtained using the hexokinase method with deproteinization (automatic analyzer).
The hexokinase method is traceable to an NIST standard.
Measuring Principle
Refer to your test strip package insert for more information.
77

Table of Contents

Other manuals for Accu-Chek Aviva Nano

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva Nano and is the answer not in the manual?

Accu-Chek Aviva Nano Specifications

General IconGeneral
Blood Sample Size0.6 µL
Test Time5 seconds
Battery LifeApproximately 1000 tests
Dimensions69 x 43 x 20 mm
DisplayLCD
Measuring Range10-600 mg/dL
Battery1 x 3V CR2032 lithium battery
Storage Conditions-25°C to 70°C
Memory500 test results
WeightApprox. 40 g (with battery)