41
Meter and Lancing Device Cleaning and Disinfecting
6
Wash hands thoroughly
with soap and water.
1
Wash hands thoroughly
with soap and water.
3
A separate Super
Sani-Cloth should be
used for cleaning and
disinfection. For
disinfecting the lancing
device, get a new cloth
and repeat step 2,
making sure that the
surface stays wet for
2minutes.
2
4
How to Clean and Disinfect the Lancing Device
WARNING
Failure to follow these instructions may damage the lancing device and stop it from working properly.
DO NOT get any moisture into any openings.
Always use the same product for both cleaning and disinfecting.
Wipe the entire surface of
the lancing device and
the inside of the cap with
a Super Sani-Cloth.