45
Maintenance and Troubleshooting
7
Display Action
Blood glucose is below the defined hypoglycemic (low blood glucose) level. See
Chapter4, Unusual Blood Glucose Results.
A general marker was assigned to this test result.
A pre-meal marker was assigned to this test result.
A post-meal marker was assigned to this test result.
A pre-meal marker was assigned to this test result and the post-meal test reminder has
been activated.
The test strip may be damaged or not properly inserted. Remove and reinsert the test
strip, or replace it if damaged.
Your blood glucose may be extremely high or a meter or a test strip error has occurred.
If your test result matches how you feel, contact your healthcare professional
immediately.
If your test result does not match how you feel, repeat the blood glucose test. See
Chapter4, Unusual Blood Glucose Results.
If the E-3code still appears for your blood glucose test, your blood glucose result
may be extremely high and above the system’s reading range. Contact your
healthcare professional immediately.
If the second test result does not match how you feel, perform a control test with
the control solution and a new test strip.
If the control result is within the acceptable range, review the proper testing
procedure and repeat the blood glucose test with a new test strip.
If the control result is not within the acceptable range, see Chapter2,
Understanding Out-of-Range Control Results.
51801_07021933001.indb 45 5/7/14 8:53 AM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva Plus and is the answer not in the manual?

Accu-Chek Aviva Plus Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
CodingNo coding required
Dimensions94 x 52 x 21 mm
Measuring Range10-600 mg/dL
DisplayLCD
Memory500 results
Battery TypeCR2032
Battery LifeApproximately 1000 tests
Automatic Power Off90 seconds after use
Weight59 g with battery