011807-EN--B (2022-12_S617 S717F S900 S950 C50_IFU)
41 / 49
Test level
(IEC 60601-1-2)
Electromagnetic environment -
comments
Immunity to proximity fields
from RF wireless
communications equipment
IEC 61000-4-3
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
Professional healthcare facility
environment and home healthcare
environment.
Electrical fast transient/burst
immunity – A.C. mains
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
± 1 kV for ports
(N/A for C50)
Professional healthcare facility
environment and home healthcare
environment.
Surge immunity
IEC 61000-4-5
± 0.5 and 1 kV
differential mode
± 0.5, 1 and 2 kV
common mode
Immunity to conducted
disturbances induced by RF
fields
(Conducted RF disturbance
immunity) – A.C. mains
IEC 61000-4-6
3 V
150 KHz at 80 MHz
6 V in ISM band and band ranging from 0.15
MHZ to 80 MHZ amateur radio band included
80 % AM at 1 kHz
Professional healthcare facility
environment and home healthcare
environment.
Power frequency magnetic
field immunity
IEC 61000-4-8
Professional healthcare facility
environment and home healthcare
environment.
Voltage dips immunity
IEC 61000-4-11
0% UT
For 0.5 cycles
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% UT
For 1 cycle
70% UT
For 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single phase at 0°
Voltage short interruptions
and voltage variations
immunity
IEC 61000-4-11
0% UT
For 250 cycles at 50
Hz
For 300 cycles at 60
Hz
Transient electrical
conduction along supply
lines
ISO 7637-2 :2011