HI Pulse or cuff The pressure in the cuff is too high
(over 299 mmHg) or the pulse is too high
(over 200 beats per minute). Relax for 5 minutes
and repeat the measurement.
LO Pulse too low The pulse is too low (less than 40 beats per
minute). Relax for 5 minutes and repeat the
measurement.
Error Messages and Troubleshooting
If an error occurs during a measurement, the measurement is interrupted, and a
corresponding error code is displayed, example <Err 3>.
Error No. Description Possible cause(s) / Solutions
Err 1 Signal too weak The pulse signals on the cuff are too weak.
Reposition the cuff and repeat the measurement.
Err 2 Error Signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat
measurement, keeping arm still.
Err 3 Abnormal cuff An adequate pressure cannot be generated in the
pressure cuff. A leak may have occurred.
Check that the cuff is correctly connected and is
not too loose. Replace the batteries, if necessary.
Repeat the measurement.
Err 5 Abnormal result The measuring signals are inaccurate, and no
result can therefore be displayed. Read through
the checklist for performing reliable measurements
on pages 25-26. Repeat the measurement.

Other manuals for ADC advantage connect 6024N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADC advantage connect 6024N and is the answer not in the manual?

ADC advantage connect 6024N Specifications

General IconGeneral
BrandADC
Modeladvantage connect 6024N
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals