Special Features
MAP (mean arterial pressure)
The e-sphyg 3 measures the patient’s true mean arterial pressure (MAP).
Each measurement includes a single MAP value. The MAP value will always be
displayed together with the systolic and diastolic blood pressure value.
Irregular heartbeat detector in «1x » Mode
The device detects irregular heartbeat in «1x» Mode. The irregular heartbeat
symbol shows up if, during the blood pressure measurement, the heart rate has
varied by more than 25%. An irregular heartbeat can affect the blood pressure
measurement. It is recommended to repeat the measurement or to choose
«Manual» Mode for verification.
MAP is determined from the maximum peak of the
oscillometric envelope curve.
22
MAP
Irregular Heartbeat
Detector