Technical Specifications
Operation temperature/humidity:
50°F to 104°F (10°C to 40°C)
15 - 90% relative maximum humidity
Storage temperature/humidity:
-4°F to 131°F (-20°C to 55°C)
15 - 90% relative maximum humidity
Weight: 2.08 lbs (942 g) (including rechargeable battery pack)
Dimensions: 7.87'' x 4.92'' x 3.54'' (20 x 12.5 x 9 cm)
Measuring method: Oscillometric
Systolic blood pressure = K1
Diastolic blood pressure = K5
Measurement range:
60 - 255 mmHg - systolic blood pressure
30 - 200 mmHg - diastolic blood pressure
40 - 200 beats per minute - pulse
Cuff pressure display:
Range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3 mmHg
Pulse accuracy: ±5% of the readout value
Power source: Rechargeable battery pack
4.8V 3500 mAh
Mains power supply DC 7.5V, 1.5 A
ADC reserves the right to alter technical specifications without prior written notice.
34
Reference to standards:
IEC 60601-1: 2005+A1:2012
IEC 60601-1-2: 2014
IEC 60601-1-6
ANSI/AAMI/ISO 81060-2
ANSI/AAMI/ISO/IEC 80601-2-30
Electromagnetic compatibility:
Device fulfills the stipulations of the standard IEC 60601-1-2.
The stipulations of the EU Directive 93/42/EEC for Medical
Devices Class IIa have been fulfilled.
0044
5000497
Type BF applied part
ETL Classified