Cleaning the Unit
Use a soft cloth with one of the following recommended cleaning solutions to
wipe the exterior of the device:
1) Mild soap and water
2) Hydrogen peroxide solution (3% diluted with water)
3) Sodium hypochlorite solution (1:10 dilution of household chloride bleach
in water)
Cleaning the Cuff
Adcuff (Two-piece design):
Remove the bladder. Fold and place the cuff cover inside a washing bag. Wash
the cuff cover with warm water (43°C; 110°F) and a mild detergent in the
washing machine.
Pasteurization: Wash the cuff cover in 75°C (167°F) hot water for 30 minutes.
Air dry the cuff. DO NOT iron the cuff cover.
Adcuff
+ (One-piece design):
Throughly saturate all surfaces of the cuff and accessories with germicidal
cleaner. Use a soft brush to remove visible soil. Rinse with water, then damp
dry.
To disinfect, throughly re-saturate all surfaces of the cuff with germicidal
cleaner. Use a soft brush to clean. Allow a five-minute wet contact time (or as
directed by the cleaner manufacturer). Do not exceed 10 minutes of wet con-
tact time.