Warning
Though the meter can display to 1499 V in Auto Range and in the 600 V
manual range, the warning triangle will be displayed above 600 V.
This indicates that the input exceeds the specified safety and operating
range. The displayed value is only indicative and not specified.
Do not continue to operate the instrument when the input value
exceeds the meter safety and operating specifications.
OL is displayed above 1499 V peak in the Peak mode and may also be
accompanied by the warning triangle if the value exceeds 600 V.
The warning triangle may also be displayed if the input signal is too
low (<50 mV).
Harmonic measurements are not operative in the DC voltage mode.
During DC voltage measurements the HARM and push buttons are
inoperative.
Analog Bar Graph
The analog bar graph continues to display voltage values while other mea-
surement functions are displayed. The bar graph ranges are 60 V and 600 V.
Overrange on the bar graph is indicated with
.
DC Voltage: Related Measurement Functions
Shift V → Peak → Ripple → V
(yellow) Push and hold (3 s) for battery life and voltage level check.
Hz Displays frequency measurements in the secondary portion
of the display.
RANGE Selects manual range of (0 - 60) or (0 - 600) volts DC.
MIN/MAX Selects record function for MIN/MAX/AVG values.
SMOOTH Displays 3-second average of measured value.
PEAK Displays maximum DC amplitude.
% Ripple Percentage AC ripple on DC signal.
Displays the highest & lowest peak values (Peak mode).