EasyManua.ls Logo

AES Dilumat S - Page 59

Default Icon
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DILUMAT S® MA108201-C
59
Adjustment Report Check Report
CALIBRATION REPORT -----------------
01 - DILUMAT S ID : XXXXXXXX (8
characters)
02 - Last calibration : XXXX/XX/XX
------------------------------------
06 - User ID : XX
07 - Date : (YYYY/MM/DD) :
XXXX/XX/XX
08 - Time : (24h) : XX:XX:XX
16 - Set weight (g) : XXXX.XX
17 - Checked weight (g) : XXXX.XX
18 - Accuracy (%) : XX.X
19 - Calibration : OK
CHECK REPORT -----------------------
01 - DILUMAT S ID : XXXXXXXX (8
characters)
02 - Last calibration : XXXX/XX/XX
------------------------------------
06 - User ID : XX
07 - Date : (YYYY/MM/DD) :
XXXX/XX/XX
08 - Time : (24h) : XX:XX:XX
16 - Set weight (g) : XXXX.XX
17 - Checked weight (g) : XXXX.XX
18 - Accuracy (%) : XX.X
20 - Check : OK
Definition of Report Data
01 Dilumat s ID Serial number for the device
02- Last calibration Date of last calibration
03- Dilution Result Result of the dilution
04- Dilution ID ID for the dilution (counter)
05- Dispensing Id ID for the dispensing (counter)
06- User ID ID for the user
07- Date (YYYY/MM/DD) Date of operation (yyyy/mm/dd format)
08- Time (24H) Time of operation
09- Dilution factor Dilution factor
10- Sample weight (g) Weight of the sample in grams
11- Dilution weight (g) Final dilution weight in grams
12- Accuracy (%) Accuracy of the dilution or dispensing
13- Sample ID
Bar code ID for the sample or sample ID (available only with the bar
code reader or keypad option)
14- Set weight (g) Weight of the dispensing set in grams
15- Dispensed weight (g) Weight of the dispensed diluent in grams
Adjustment and Check Report Data
16- Set weight (g) Weight of the calibration set in grams
17- Checked weight (g) Actual weight of the calibration in grams
18- Accuracy (%) Accuracy of the calibration or verification
19- Calibration Conclusion of the calibration
20- Check Conclusion of the verification
Simulation Report Data
21- Set cycles Number of test cycles programmed
22- Completed cycles Number of test cycles completed
23- Counter 1 Number of dilutions completed since the last reset
24- Counter 2 Number of dispensing operations completed since the last reset
25- Counter 3 Number of hours and minutes powered on
26- Status Conclusion of the test phase
Note:
- These reports are published only in English.
- KO indicates a failure.

Table of Contents