As in the Full Auto mode, End Locate mode offers automatic OTDR setup but replaces the Event
Table view with the End Locate view to indicate the fiber end location in user selected units of
measure. In the End Locate mode, the End Locate view is always shown after a test is started.
In the Live test mode, you may set to Off or By Range.
• Off: Setting the Auto Setup parameter to Off, allows you to set the Range, Filter, and Pulse
parameters while the Averaging parameter is set by OFL250.
• By Range: Setting the Auto Setup option to By Range, allows you to set the Range
parameter while the Filter, Pulse and Averaging parameters are set by OFL250.
To Set Live Mode Parameters
1. From the displayed Live mode submenu, use keys to highlight the desired parameter.
2. Use keys to change the highlighted parameter.
OTDR Test Settings
End Locate Mode Settings
Live Mode Settings
Summary Tab (dB icon) only
present if ORL upgrade
option is installed.
CABLE000_001
1550
ZOOM
UNZOOM
End: X.X m
dB
i
End location in user
selected units.
End Locate view tab.
Trace Graph.
Set the desired test wavelength.
Use Left/Right arrow key to
change highlighted parameter.
If set to By Range, set Range
while the OFL250 sets the
Filter, Pulse, and Averaging.
If set to Off, set Range, Filter,
Pulse, and Averaging.