IMPORTANT: Only use AgaMatrix WaveSense JAZZ Test
Strips with the AgaMatrix WaveSense JAZZ System.
Important Test Strip Information:
[1] Store the AgaMatrix WaveSense JAZZ Test Strip vial in a
cool, dry place at 8 °C to 30 °C (46 °F to 86 °F).
[2] Use AgaMatrix WaveSense JAZZ Test Strips only within the
system operating temperature range of 10 °C to 40 °C (50 °F to
104 °F).
[3] Keep away from direct sunlight and heat.
[4] Store your AgaMatrix WaveSense JAZZ Test Strips in their
original vial only; never store them in another vial, any other
container, or outside the vial.
[5] After removing an AgaMatrix WaveSense JAZZ Test Strip
from the vial, immediately close the vial cap tightly.
[6] With clean, dry hands you may gently handle the
AgaMatrix WaveSense JAZZ Test Strip when removing it from
the vial or inserting it into the meter.
[7] Do not use AgaMatrix WaveSense JAZZ Test Strips beyond
the expiration date or 90 days after first opening the vial. This
may cause inaccurate results. Write the discard date (90 days
from the first opening) on the test strip vial.
[8] Do not bend, cut or alter AgaMatrix WaveSense JAZZ Test
Strips.