EasyManua.ls Logo

AGFA XD 17 - Page 134

AGFA XD 17
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
100% reduc-
tion for
250/300 cycles
(5 sec.)
100% reduc-
tion for
250/300 cy-
cles (5 sec.)
terruptions or a bat-
tery.
Magnetic field at
the supply frequen-
cy (50/60 Hz) in ac-
cordance with IEC
61000-4-8
30 A/m 30 A/m Magnetic field at
the network fre-
quency should cor-
respond to the typi-
cal values as they
are in a commercial
and clinical envi-
ronment.
Tests of Resistance to
Disruption
IEC 60601
Test Level
Level of
Agreement
Electromagnetic En-
vironment
Conducted high fre-
quency disturbance
variables in accord-
ance with
IEC 61000-4-6
3 V
eff
150
kHz to 80
MHz
6 V
eff
in the
ISM bands
150 KHz to
80 MHz
3 V
eff
150
kHz to 80
MHz
6 V
eff
in the
ISM bands
150 KHz to
80 MHz
The electromagnetic
field strength of a
stationary RF trans-
mitter determined by
an electromagnetic
test survey must be
less than the compli-
ance level of each
frequency range.
Radiated high fre-
quency disturbance
variables in accord-
ance with
IEC 61000-4-3
3 V/m
80 MHz to
2.7 GHz
3 V/m
80 MHz to
2.7 GHz
Interference may oc-
cur in the vicinity of
equipment marked
with the symbol:
Note: The higher value will apply at 80 MHz and 800 MHz.
134 | XD 17, XD17 | Remarks for HF-emission and immunity
0434A EN 20210426 1355

Table of Contents

Related product manuals