Use of protective plastic bag
WARNING:
Liquids ingressing the DR Detector may cause malfunction and
contamination.
If there is a chance that the detector comes in contact with liquids (bodily
fluids, disinfectants,...), the DR Detector must be wrapped in a protective
plastic bag while performing the examination.
It is considered good clinical practice to use a single-use protective bag in all
cases where contact of the device or contaminants is expected, to avoid
contamination of others.
Make sure that the plastic bag is not wrinkled to avoid the creases showing in
the image.