Install Long Life Pump Heads (Quaternary and Flexible Pumps)
Install the Pump Head Assembly
3
Insert the inlet weaver to the inlet valve (1). Fix the weaver
with the plastic screw to the inlet valve (2).
4
Connect the capillary from the outlet adapter on the secondary
pump head to the pressure sensor.
5
Connect the seal wash tubes.
6
Close the doors.
7
Perform a Pump Leak Rate Test.

