Upgrade the Valve Support (G4204A only)
Secondary Pump Head Parts
*01200-90121*
*01200-90121*
01200-90121
Part Number: 01200-90121
Rev. B
Edition: 11/2016
Printed in Germany
©
Agilent Technologies, Inc 2016
Agilent Technologies, Inc
Hewlett-Packard-Strasse 8
76337 Waldbronn
Germany
Upgrade the Valve Support (G4204A only)
Tools required p/n Description
G7120-68708 HPLC System Tool Kit-Infinity-II
Parts required p/n Description
G4204-04000 Valve Support
Preparations Remove capillaries:
from the pump head to the pressure sensor,
from the pressure sensor to the multipurpose valve,
from valve port 4.
Remove the pressure sensor.
Remove the Multipurpose valve.
Remove the inline filter (if installed).
Remove the Jet Weaver (if installed).
Remove the valve support as described in the Service Manual.
1
Insert the valve drive into the new valve support delivered with
the upgrade kit.
2
Fix the valve drive in the new valve support.
3
Slide the valve support into the pump (1.) and fix it with screw
(2.).
4
Install the pressure sensor, the multipurpose valve, and
(optionally) the Jet Weaver.
Next Steps:
5
Reconnect all capillaries.
6
Install the inline filter.



Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals