Remove Pump Heads (Binary and High Speed Pumps)
Remove the Pump Head Assembly
5
Remove the capillary connection at the top of the secondary
pump head to the pump valve.
6
Remove the flow connection between the degassing unit and
the primary pump head inlet.
7
Loosen the inlet valve. Keep the inlet valve installed to the
pump head assembly.
8
Counter the lock screw of the heat exchanger capillary while
loosening the outlet valve. Keep the outlet valve installed to
the pump head assembly.
9
Loosen the high pressure filter. Keep the filter installed to the
pump head assembly.
10
Open the four screws holding the pump heads.
Open the screws step by step, not screw by screw.