Manual Change
Source Voltage Accuracy Test
Step 16. Calculate the test result according to the calculation sheet, and record the result into the
performance test record.
Current Measurement Accuracy Test
The 4339B measures the current through the calibrated 16353A/16353H , and the
measured values are compared with the current values calculated from the
16353A/16353H’s calibration value and the voltage across the standard.
Specification
Test Equipment
Procedure
Step 1. Record the 16353A/16353H Standard Resistor Set calibration values on the calculation
sheet.
Step 2. Reset the 4339B using the following procedure:
1. Press [blue] [] to display the system reset menu.
2. Select Yes using [][] or [] [] and press [Enter].
Step 3. Set the Offset-error Canceling to ON using the following procedure:
1. Press [blue][-] to display the configuration setting menu.
2. Select Offset using [][] or [][] and press [Enter].
3. Select On using [][] or [][] and press [Enter].
4. Select Exit and press [Enter] to exit the configuration setting menu.
Step 4. Set up the equipment as shown in Figure 4.
Basic Measurement Accuracy: ± 0.4% (See Chapter 8 General Information for
details.)
Description Recommended Model
Multimeter 3458A
Standard Resistor Set 16353A, 16353H
RC Box Adapter PN 04339-65005
Dual Banana-BNC(m) Cable PN 11001-60001