Manual Change
Source Voltage Accuracy Test
Figure 4 Current Measurement Accuracy Test Setup
Step 5. Set the RC Box Adapter switch to the F position.
Step 6. Press the 3458A Multimeters [DCV] to set the measurement mode to DC voltage.
Step 7. Press [blue] [5]. Select
ExecCal and press [Enter] to perform the calibration.
Step 8. Set the source voltage to 10V using the following procedure:
1. Press [Source Voltage] to display the source voltage setup menu.
2. Press [1][0][Enter] to set the source voltage to 10V.
3. Confirm that +10V is displayed on the right side of the LCD.
Step 9. Press [V Output] to turn the source voltage ON. (V Output indicator turns ON.)
Step 10. Press [blue] [4]. Select
OpenMeas and press [Enter] to perform the OPEN correction.
Step 11. Press [V Output] to turn the source voltage OFF. (V Output indicator turns OFF.)
Step 12. Connect the 16353B to the RC Box Adapter as shown in Figure 5.
4339boe1006
4339B
To Monitor H
Dual Banana-BNC(m) Cable
: No Connection
RC Box
Adapter
Multimeter
