Checkout Information 2
Natural Gas Analyzers 23
Analysis of the Universal Gas Calibration standard results in a
chromatogram similar to Figure 25.
CAUTION
The CP-Molsieve 5A channel is factory tuned on carrier gas helium.
If analysis of helium or hydrogen is required on the CP-Molsieve 5A
channel, the carrier gas must be changed and configured for argon.
The procedure for configuring the carrier gas in OpenLAB CDS
EZChrom is described in Appendix D on page 31.
Figure 25 Universal Gas Calibration standard analyzed with the CP-Molsieve 5A channel on Analyzer B
Extended
Seconds
20 30 40 50 60 70 80 90 100 110
mV
0
2
4
6
8
Helium
Neon
Hydrogen
Oxygen
Nitrogen
Methane
Carbon Monoxide