EasyManua.ls Logo

Agilent Technologies 490 Micro GC

Agilent Technologies 490 Micro GC
37 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix C: Typical Method Settings for Natural Gas Analyzers
Natural Gas Analyzers 30
Table 3 Method for Agilent 490 Micro GC - Natural Gas Analyzer B
Method settings
*
* For more details of the method, see the pdf-method file available on the Natural Gas
Analyzer CD.
PoraPLOT U CP-Sil 5 CB_6m
Carrier gas Helium Helium
Injector temperature (°C) 110 110
Injection time (ms) 40 40
Backflush time (s)
The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U,
4 meter CP-Sil 5 CB, and HayeSep A channel.
17 -
Column temperature (°C) 60 70
Pressure (kPa) 175 150
Sample line temperature (°C) 110 110
Table 4 Method for Agilent 490 Micro GC - Natural Gas Analyzer B Extended
Method settings
*
CP-Molsieve 5A PoraPLOT U CP-Sil 5 CB_6m
Carrier gas Helium Helium Helium
Injector temperature (°C) 110 110 110
Injection time (ms) 40 40 40
Backflush time (s)
11 17 -
Column temperature (°C) 80 60 70
Pressure (kPa) 200 175 150
Sample line temperature (°C) 110 110 110
* For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD.
The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U, 4 meter CP-Sil 5 CB, and HayeSep A channel.

Other manuals for Agilent Technologies 490 Micro GC

Related product manuals