214 53220A/53230A User’s Guide
6 53220A/53230A Math, Graphing, and Data Logging
Figure 6-7. 15-Bin (Point) Histogram.
Histogram computation is enabled with the command:
CALCulate2:TRANsform:HISTogram[:STATe] {OFF|ON}
CALCulate2:TRANsform:HISTogram[:STATe]? (query form)
ON - enables histogram computation.
OFF - disables histogram computation. Following a reset (*RST), instrument
preset (SYSTem:PRESet), or front panel (Preset) histograms are disabled.
CALCulate2:TRANsform:HISTogram:STATe ON
CALCulate2:TRANsform:HISTogram:RANGe:LOWer
CALCulate2:TRANsform:HISTogram:RANGe:UPPer
C
ALCulate2:TRANsform:HISTogram:POINts
CALCulate2:TRANsform:HISTogram:RANGe:AUTO
CALCulate2:TRANsform:HISTogram:RANGe:AUTO:COUNt
CALCulate2:TRANsform:HISTogram:CLEa
r
CALCulate2:TRANsform:HISTogram:COUNt?