53220A/53230A User’s Guide 215
53220A/53230A Math, Graphing, and Data Logging 6
The number of readings represented by a histogram (Figure 6-7) can be que-
ried with the command:
CALCulate2:TRANsform:HISTogram:COUNt?
Setting Up the Histogram
The counter builds a histogram based on number of bins (points), a lower
range, and an upper range.
Number of Bins
The number of bins is specified with the command:
CALCulate2:TRANsform:HISTogram:POINts {<value>|MINimum|
MAXimum|DEFault}
CALCulate2:TRANsform:HISTogram:POINts? [{MINimum|
MAXimum|DEFault}] (query form)
value - specifies the number of bins (points) between the lower and upper
range values. value can range from 10 to 1,000. Following a reset (*RST) or
instrument preset (SYST:PRESet) the number of bins is set to 100.