94 5975/5977 Series MSD Operation Manual
4 General Maintenance
To Open the Analyzer Chamber
The analyzer chamber should only be opened to clean or replace the ion
source, change the detector’s EM, or to change a filament.
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Wrist strap, antistatic
Small (9300-0969)
Medium (9300-1257)
Large (9300-0970)
Procedure
1 Vent the MS. (See “Pumping Down” on page 65.)
2 Open the left side panel. (See “To Open the Analyzer Chamber” on
page 94.)
3 Loosen the front analyzer side plate thumbscrews (Figure 10) if they are
fastened.
CAUTION
Electrostatic discharges to analyzer components are conducted to the quad driver
board, where they can damage sensitive components. Wear a grounded antistatic
wrist strap and take other antistatic precautions (see “Electrostatic discharge” on
page 90) before you open the analyzer chamber.
WARNING
The analyzer, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.

Table of Contents

Other manuals for Agilent Technologies 5977 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977 MSD and is the answer not in the manual?

Agilent Technologies 5977 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977 MSD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals