EasyManuals Logo

Agilent Technologies 6495B User Manual

Agilent Technologies 6495B
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
3 Installation Completion and System Verification
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
68 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Figure 35 Example of a Negative mode signal-to-noise report.
6 Change solvents (to A = water w/0.1% formic acid and B = acetonitrile for
Positive Ion Polarity).
7 Clean the injectors. See step 12 on page 52.
8 Equilibrate the column for 1 hour at 30:70 A:B.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6495B and is the answer not in the manual?

Agilent Technologies 6495B Specifications

General IconGeneral
BrandAgilent Technologies
Model6495B
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals