3 Installation Completion and System Verification
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
70 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Tabl e 5 Installation Sensitivity Performance Specifications
Model Mode G1948B Electrospray Agilent Jet Stream
6495 Positive/Negative 75,000:1 (Auto-RMS,1pg)
6490 Positive/Negative 25,000:1 (Auto-RMS,1pg)
6470 Positive 37,500:1 (Auto-RMS,1pg)
Negative 15,000:1 (Auto-RMS,1pg)
6460 Positive 5,000:1 (Auto-RMS,1 pg) 15,000:1 (Auto-RMS,1 pg)
Negative 1,500:1 (Auto-RMS,1 pg) 5,000:1 (Auto-RMS,1 pg)
6430 Positive 300:1 (1×RMS, 1 pg)
6420 Positive 3,000:1 (Auto-RMS,1 pg)
Negative 1,000:1 (Auto-RMS,1 pg)
6410B
with
G2573A
Positive 300:1 (1×RMS, 1 pg)
6410B Positive 150:1 (1×RMS, 1 pg)
Tabl e 6 G6061A/G6062A/G6063A Marketing Bid Sensitivity Performance
Specifications
Model Mode G1948B Electrospray Agilent Jet Stream
6495 Positive/Negative 150,000:1 (Auto-RMS,1pg)
6490 Positive/Negative 50,000:1 (Auto-RMS,1pg)
6470 Positive (G6061A) 75,000:1 (Auto-RMS,1pg)
Negative (G6062A) 30,000:1 (Auto-RMS,1pg)
6460 Positive (G6061A) 10,000:1 (Auto-RMS, 1 pg) 30,000:1 (Auto-RMS,1 pg)
Negative (G6062A) 3,000:1 (Auto-RMS, 1 pg) 10,000:1 (Auto-RMS,1 pg)
6420 Positive (G6061A) 6,000:1 (Auto-RMS, 1 pg)
Negative (G6062A) 2,000:1 (Auto-RMS, 1 pg)